REPEAT ANALYSIS CITRA RADIOGRAF PADA PESAWAT DIGITAL RADIOGRAPHY (DR) DI INSTALASI RADIOLOGI RSD K.R.M.T. WONGSONEGORO KOTA SEMARANG

Authors

  • Adit Khoikil Anwar Universitas `Aisyiyah Yogyakarta
  • Arnefia Mei Yusnida Universitas `Aisyiyah Yogyakarta
  • Muhamad Fa'ik Universitas `Aisyiyah Yogyakarta

DOI:

https://doi.org/10.51878/healthy.v5i4.15667

Keywords:

Repeat Radiograf, Radiografi Konvensional, Faktor Penyebab, Quality Assurance/Quality Control (QA/QC)

Abstract

ABSTRACT

Repeat analysis is part of radiological quality assurance (QA) aimed at identifying and analyzing radiographic images that require repetition and determining their causes. This activity is important for improving service quality, reducing unnecessary repeat examinations, and minimizing additional radiation exposure to patients. This study aimed to determine the percentage of repeat examinations, identify the factors causing repeats, and describe the efforts made to minimize repeat occurrences at the Radiology Department of RSD K.R.M.T. Wongsonegoro, Semarang City. This study employed a descriptive quantitative method with an observational approach and was conducted from October 2025 to June 2026. Data were collected through observation, documentation, and interviews and were analyzed descriptively. During January–March 2026, 109 repeat cases were identified, consisting of patient motion in 53 cases (48.6%), artifacts in 41 cases (37.6%), and positioning errors in 15 cases (13.8%). Efforts to minimize repeats included improving radiographers’ competencies, optimizing patient communication and education, applying positioning according to established procedures, controlling artifacts, implementing Quality Assurance (QA) and Quality Control (QC), performing equipment calibration, and conducting regular reject-repeat analysis. Patient motion was the most dominant factor contributing to repeat examinations; therefore, continuous control of patient, radiographer, and equipment-related factors is necessary to maintain image quality and reduce repeat examinations.

ABSTRAK

Repeat analysis merupakan bagian dari quality assurance (QA) radiologi untuk mengidentifikasi dan menganalisis citra radiograf yang perlu diulang serta penyebabnya. Kegiatan ini penting untuk meningkatkan kualitas pelayanan, mengurangi pengulangan pemeriksaan, dan meminimalkan paparan radiasi tambahan pada pasien. Penelitian ini bertujuan mengetahui persentase repeat, faktor penyebab, dan upaya meminimalkan kejadian repeat di Instalasi Radiologi RSD K.R.M.T. Wongsonegoro Kota Semarang. Penelitian menggunakan metode kuantitatif deskriptif dengan pendekatan observasional yang dilaksanakan pada Oktober 2025 hingga Juni 2026. Data dikumpulkan melalui observasi, dokumentasi, dan wawancara, kemudian dianalisis secara deskriptif. Selama Januari–Maret 2026 ditemukan 109 kasus repeat, yang terdiri atas pergerakan pasien sebanyak 53 kasus (48,6%), artefak 41 kasus (37,6%), dan kesalahan posisi 15 kasus (13,8%). Upaya meminimalkan repeat dilakukan melalui peningkatan kompetensi radiografer, optimalisasi komunikasi dan edukasi pasien, penerapan positioning sesuai prosedur, pengendalian artefak, pelaksanaan Quality Assurance (QA) dan Quality Control (QC), kalibrasi peralatan, serta reject-repeat analysis secara berkala. Pergerakan pasien merupakan faktor penyebab repeat yang paling dominan, sehingga pengendalian faktor pasien, radiografer, dan peralatan perlu dilakukan secara berkelanjutan untuk mempertahankan kualitas citra dan mengurangi pengulangan pemeriksaan.

References

Alashban, Y., Shubayr, N., Alghamdi, A. A., Alghamdi, S. A., & Boughattas, S. (2022). An assessment of image reject rates for digital radiography in Saudi Arabia: A cross-sectional study. Journal of Radiation Research and Applied Sciences, 15(1), 219–223. https://doi.org/10.1016/j.jrras.2022.01.023

Armitasari, L., Mubaroq, R., & Putro, W. M. C. (2024). Analisis repeat citra radiograf sebagai bentuk keselamatan radiasi di Instalasi Radiologi Rumah Sakit Islam Klaten. Prosiding Seminar Si-INTAN, 4(1), 18–24. https://doi.org/10.53862/SSI.v4.092024.003

Artitin, C., Syafputri, W. M., & Nansih, L. A. (2022). Analisis pengulangan (repeat) radiograf di unit radiologi Rumah Sakit Umum Daerah Mohammad Natsir Solok. Jurnal Teras Kesehatan, 5(2), 25–30. https://jurnal.politeknikalislam.ac.id/index.php/Jutek/article/download/72/57/934

Bontrager, K. L., & Lampignano, J. P. (2018). Bontrager's textbook of radiographic positioning and related anatomy (9th ed.). Elsevier.

Chafidhi, A., Suraningsih, N., & Budiwati, T. (2018). Analisis pengulangan citra digital dengan menggunakan digital radiography di instalasi radiologi Rumah Sakit Panti Waluyo Surakarta. RadX: Jurnal Ilmiah Radiologi, 3(2). https://journalwh.uwhs.ac.id/index.php/jdx/article/view/22

Dave, J. K., Jones, A. K., Fisher, R., Hulme, K., Rill, L., Zamora, D., Woodward, A., Brady, S., MacDougall, R. D., Goldman, L., Lang, S., Peck, D., Apgar, B., Shepard, S. J., Uzenoff, R., & Willis, C. (2018). Current state of practice regarding digital radiography exposure indicators and deviation indices: Report of AAPM Imaging Physics Committee Task Group 232. Medical Physics, 45(11), e1146–e1160. https://doi.org/10.1002/mp.13212

Falentin, F. F., Astari, F. M., & Za’im, M. (2025). Identifikasi faktor penyebab pengulangan citra radiografi sebagai upaya menjaga keselamatan pasien di Instalasi Radiologi Rumah Sakit Umum ‘Aisyiyah Ponorogo. Journal of Innovation Research and Knowledge, 5(4), 4481–4496. https://doi.org/10.53625/jirk.v5i4.11228

Firmansyah. (2020). Pemberian terapi latihan pada kasus post ORIF fracture femur 1/3 distal sinistra di RSUD Kota Semarang. Research of Service Administration Health and Sains Healthys, 1(1), 25–32. https://doi.org/10.58258/rehat.v1i1.3163

Fitri, A. W., Mahanani, A., & Fakhrurreza, M. (2025). Analisis pengulangan (repeat) radiograf computed radiography di Instalasi Radiologi Rumah Sakit Bhakti Wira Tamtama Semarang. Jurnal Kesehatan Tambusai, 6(2), 7340–7346. https://doi.org/10.31004/jkt.v6i2.45089

Hasaneen, M., AlHameli, N., AlMinhali, A., Alshehhi, S., Salih, S., & Alomaim, M. M. (2023). Assessment of image rejection in digital radiography. Journal of Medicine and Life, 16(5), 731–735. https://doi.org/10.25122/jml-2022-0341

International Atomic Energy Agency. (2023). Handbook of basic quality control tests for diagnostic radiology (IAEA Human Health Series No. 47). International Atomic Energy Agency. https://doi.org/10.61092/iaea.rhjo-8d0u

Jones, A. K., Heintz, P., Geiser, W., Goldman, L., Jerjian, K., Martin, M., Peck, D., Pfeiffer, D., Ranger, N., & Yorkston, J. (2015). Ongoing quality control in digital radiography: Report of AAPM Imaging Physics Committee Task Group 151. Medical Physics, 42(11), 6658–6670. https://doi.org/10.1118/1.4932623

Kementerian Kesehatan Republik Indonesia. (2008). Keputusan Menteri Kesehatan Republik Indonesia Nomor 129/Menkes/SK/II/2008 tentang standar pelayanan minimal rumah sakit. https://studylibid.com/doc/4289320/kepmenkes-no.129-tahun-2008-standar-pelayanan-minimal-rs

Kementerian Kesehatan Republik Indonesia. (2020). Peraturan Menteri Kesehatan Republik Indonesia Nomor 24 Tahun 2020 tentang pelayanan radiologi klinik. Berita Negara Republik Indonesia Tahun 2020 Nomor 1058. https://jdih.kemkes.go.id/documents/peraturan-menteri-kesehatan-nomor-24-tahun-2020

Kementerian Kesehatan Republik Indonesia. (2026). Peraturan Menteri Kesehatan Republik Indonesia Nomor 6 Tahun 2026 tentang Rumah Sakit. Berita Negara Republik Indonesia Tahun 2026 Nomor 382. https://jdih.kemkes.go.id/documents/peraturan-menteri-kesehatan-nomor-6-tahun-2026

Little, K. J., Reiser, I. S., Apgar, B., Dalal, P., Dave, J. K., Fisher, R. F., Hulme, K. W., Jafari, M. E., Marshall, E. L., Meyer, S. W., Moore, Q. T., Murphy, N. E., Nishino, T. K., Nye, K., O'Donnell, K., Sabol, J. M., Sanchez, A. A., Sensakovic, W. F., Tarbox, L. R., Uzenoff, R. A., Walz-Flannigan, A. I., Willis, C. E., & Zhang, J. (2023). AAPM task group report 305: Guidance for standardization of vendor-neutral reject analysis in radiography. Journal of Applied Clinical Medical Physics, 24(5), e13938. https://doi.org/10.1002/acm2.13938

Misah, A., & Aris Diartama, A. A. (2025). Analisis pengulangan (repeat) eksposure pada modalitas digital radiography (DR) di Rumah Sakit Daerah (RSD) Mangusada Badung. Jurnal Surya Medika (JSM), 11(2), 99–105. https://doi.org/10.33084/jsm.v11i2.7716

Rochmayanti, D., Adi, K., & Edi Widodo, C. (2023). Repeat analysis program as a quality assurance system for radiology management: Causal repeat and challenges. E3S Web of Conferences, 448, 05004. https://doi.org/10.1051/e3sconf/202344805004

Sabilonnaja, M. P., Liscyaningsih, I. A. N., & Utami, A. P. (2024). Analisis pengulangan citra digital dengan menggunakan digital radiography di instalasi radiologi RSUD dr. R. Soetijono Blora. Prosiding Seminar Nasional Penelitian dan Pengabdian Kepada Masyarakat LPPM Universitas ’Aisyiyah Yogyakarta, 2, 2130–2135. https://proceeding.unisayogya.ac.id/index.php/prosemnaslppm/article/view/203

Saifudin, Felayani, F., & Al Amin, K. (2023). Pendampingan program kendali mutu pelayanan radio diagnostik non rumah sakit di Kabupaten Banyumas. Segantang Lada: Jurnal Pengabdian Kesehatan, 1(1), 6–12. https://doi.org/10.53579/segantang.v1i1.98

Sari, A. N. V., Yusnida, A. M., & Fakhrurreza, M. (2025). Analisis pengulangan foto rontgen pada modalitas digital radiography di Instalasi Radiologi di RSUD Muntilan. PREPOTIF: Jurnal Kesehatan Masyarakat, 9(2), 6737–6745. https://doi.org/10.31004/prepotif.v9i2.49081

Seibert, J. A., & Morin, R. L. (2011). The standardized exposure index for digital radiography: An opportunity for optimization of radiation dose to the pediatric population. Pediatric Radiology, 41(5), 573–581. https://doi.org/10.1007/s00247-010-1954-6

Shepard, S. J., Wang, J., Flynn, M., Gingold, E., Goldman, L., Krugh, K., Leong, D. L., Mah, E., Ogden, K., Peck, D., Samei, E., Wang, J., & Willis, C. E. (2009). An exposure indicator for digital radiography: AAPM Task Group 116 (executive summary). Medical Physics, 36(7), 2898–2914. https://doi.org/10.1118/1.3121505

Suharmono, B. H., Anggraini, I. Y., Hilmaniyya, H., & Astuti, S. D. (2020). Quality assurance (QA) dan quality control (QC) pada instrumen radioterapi pesawat LINAC. Jurnal Biosains Pascasarjana, 22(2), 73–80. https://doi.org/10.20473/jbp.v22i2.2020.73-80

Wakhusna, A. N., Liscyaningsih, I. A. N., & Za’im, M. (2023). Studi kasus pengulangan foto rontgen pasien pediatrik pada modalitas computed radiography (CR) di RSUD Dr. R. Soetijono Blora [Tugas akhir, Universitas ‘Aisyiyah Yogyakarta]. http://digilib.unisayogya.ac.id/id/eprint/6854

Wijaya, I. M., & Arianti, N. W. M. (2025). Repeat analysis citra modalitas digital radiography merk United Imaging di Unit Radiologi RSU Prima Medika Denpasar: Repeat analysis of digital radiography images United Imaging modality at the Radiology Unit of RSU Prima Medika Denpasar. Jurnal Surya Medika (JSM), 11(1), 246–252. https://journal.umpr.ac.id/index.php/jsm/article/view/7534

Yoga, A. A. N. G. K., Astina, K. Y., & Juliasa, I. W. (2025). Analisis daily quality assurance (QA) / quality control (QC) pada pesawat Linac di Unit Onkologi Radiasi RSUD Bali Mandara. Kappa Journal, 9(1), 35–42. https://doi.org/10.29408/kpj.v9i1.27443

Yunita, Y., Mahani, A., & Za’im, M. (2025). Studi kasus repeat (pengulangan) radiograf dengan modalitas digital radiography (DR) di Instalasi Radiologi RSUD Dr. Soehadi Prijonegoro Sragen. Jurnal Ilmiah Kedokteran dan Kesehatan, 4(2), 227–241. https://doi.org/10.55606/klinik.v4i2.3929

Downloads

Published

2026-09-30

How to Cite

Anwar, A. K., Yusnida , A. M., & Fa’ik, M. (2026). REPEAT ANALYSIS CITRA RADIOGRAF PADA PESAWAT DIGITAL RADIOGRAPHY (DR) DI INSTALASI RADIOLOGI RSD K.R.M.T. WONGSONEGORO KOTA SEMARANG. HEALTHY : Jurnal Inovasi Riset Ilmu Kesehatan, 5(4), 1513–1529. https://doi.org/10.51878/healthy.v5i4.15667